| DC Resistance Measurement |
| Range |
0.1 M Ω to 20 K Ω (four-wire measurement) |
| Basic Accuracy |
0.1 M Ω to 200 MΩ ± (0.1 % ± 0.5 m Ω) 200 m Ω or 20 k Ω ± 0.1 % |
| Inductance Measurement |
| Test Parameter |
Ls, Lp |
| Test Frequency |
50 Hz, 60 Hz, 100 Hz, 120 Hz, 1 kHz, 10 kHz, 20 kHz, 40 kHz, 50 kHz, 100 kHz |
| Test Level |
0.1 V, 0.3 V, 1.0 V |
| Output Impedance |
100 Ω |
| Basic Accuracy |
0.1 % |
| Measurement Display Range |
0.001 uH to 9.9999 kH |
| Insulation Resistance Test |
| Out-Out Voltage |
0.100 kV to 1.000 kV ± (1.5 % of reading + 5 counts) |
| Resistance Measurement Range |
1 M Ω to 10 G Ω |
| Test Time |
0.3 s to 999.9 s |
| Accuracy |
< 500 V 1 MΩ to 1 GΩ: ± (10 % of reading + 5 counts) 500 V 1 MΩ to 1 GΩ: ± (5 % of reading + 5 counts) 1 GΩ to 10 GΩ: ± (10 % of reading + 5 counts) |
| Ac/Dc Withstanding |
| Output-Voltage Range |
AC: 0.050 kV to 5.000 kV (± 2 % of reading + 5 digits) DC: 0.050 kV to 6.000 kV (± 2 % of reading +5 digits) |
| Current Measurement Range |
AC: 030 mADC: 010 mA |
| Arc Detect |
AC: 1 mA 15 mADC: 1 mA 10 mA |
| Test Time |
0.3 s to 999.9 s |
| Accuracy |
± 2 % |
| Impulse Winding Test |
| Impulse Voltage |
100 V to 5000 V, 10 V step, ± 5 % ± 10 V |
| Comparison Methods |
Area, differential area, corona and differential phase |
| Test Repeated Accuracy |
1 % |
| Other Functions |
| Test Channels |
4 to 12 |
| Test Statistics |
Statistics function of measurement time and results, up to 1000 pieces |
| Memory |
Internal 180 groups (standard waveform data and test setup parameters) U-disk 600 groups |
| Interface |
HANDLER (START, STOP, PASS, FAIL, EOM) USB Device, RS232C USB Host (support storage of BMP, GIF, PNG image files and firmware upgrade) |
| Temperature & Humidity |
32 °F to 104 °F (0 °C to 40 °C) and 90 % RH |
| Power Requirements |
Voltage 99 V to 121 V or 198 V to 242 V |
| Frequency 47.5 Hz to 63 Hz |
| Power |
≤ 300 VA |
| Dimension (W × H × D) |
16.92 in x 6.88 in x 17.32 in (430 mm × 175 mm × 440 mm) |