Emissions And Stack Digital Bridges LCR Generator

Emissions And Stack digital bridges LCR generators are high-precision component parameter analyzers that make use of the automatic balancing bridge principle.

SKU: EMSLCR-295 Category:

Overview

Emissions And Stack digital bridges LCR generators are high-precision component parameter analyzers that make use of the automatic balancing bridge principle. They feature a testing bandwidth of 10 Hz to 1 MHz, continuously adjustable frequencies, 0.05% of basic measuring accuracy, automatic level control, list scanning, and position counting. These generators can provide accurate and comprehensive measurements and analyses for most components and materials, so they are widely used in product R&D, component incoming inspections, and online product sorting. 

 

Features

  • 0.05% basic accuracy 
  • A maximum of 200 time/s measuring speed 
  • Frequency measurement range of 10 Hz to 1 MHz, Continuous adjustable, 1 mHZ step by step 
  • Amplitude of excitation signal 10 mV to 2 V adjustable,1 mV step by step 
  • Internal programmable DC bias voltage -2 V to +2 V 
  • Support external dc bias voltage applied -60 V to +60 V 
  • Support external current source 
  • V,I testing signal monitoring 
  • 10-point list scanning testing 
  • 10 grade sorting and counting  
  • 100 groups of correcting data for specified frequencies 
  • Automatic range, manual range 
  • 7 in (177.8 mm) LED display screen, English interfaces 
  • USB, Ethernet, RS232, GPIB, Handler interfaces 

 

 

Technical Specifications

Power voltage 220 V AC or 50 Hz ±10%, 
Power consumption < 20 W 
Display 7 in (177.8 mm) TFT LCD, resolution 800 pixels x 480 pixels 
Interfaces LAN, RS232, GPIB, USB Host, USB Device and Handler 
Service environment 32 °F to 104 °F (0 °C to 40 °C) 
Sizes 12.9 in x 11.22 in x 5.35 in (330 mm x 285 mm x 136 mm) 
Weight  7.93 lb (3.6 kg) 
Test signal frequency range 10 Hz to100 kHz 
Frequency resolution, accuracy Resolution: 1 mHz, Accuracy: 0.01% 
Test parameters Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs,
Lp-D, Lp-Q, Lp-G, Lp-Rp, Ls-D, Ls-Q, Ls-Rs,
Rs-Xs, |Z|-θr, |Z|-θd, |Y|-θr, | Y|-θd, G-B 
Measuring display speed Fast: 50 times/s (20 ms), Medium:10 times/s (100 ms), Slow: 1.25 times/s (800 ms) 
Custom measuring speed Between 0.5 times/s to 200 times/s, can be set 
LCR parameter range Cp, Cs: 0.001000 pF to 99.9999 F
Lp, Ls: 0.001000 nH to 99.9999 kH
Rp, Rs, |Z|, Xs: 0.001000 mΩ to 999.999 MΩ
G, B, |Y|: 0.001000 μS to 999.999 kS
θr: ± 0.000001rad to 3.14159 rad
θd: ± 0.000001 deg to 179.9999 deg
D: ± 0.000001 to 9.99999
Q: ± 0.001 to 99999.9 
Test signal voltage range 0 Vrms to 2 Vrms 
Voltage resolution, accuracy Resolution: 1 mV, accuracy: 5% +5 mV 
Test signal current range 100 μArms to 20 mArms 
Current resolution, accuracy Resolution: 10 μA, accuracy: 5% +50 μA 
DC bias voltage source Internal: -2 V to +2 V voltage bias, -20 mAto +20 mA current bias
External: -60 V to +60 V voltage bias 
Internal resistance of signal source 30 Ohm or 100 Ohm, selectable 
Basic accuracy 0.05% 
Display resolution 6 1/2 digit 
Comparator 9 groups of qualified setting, one group of unqualified setting, one group of auxiliary setting 
Triggered mode Internal, External, manual, bus 
Mathematical operations Delta (absolute value), Delta% (percentage), Direct reading 
Calibration function Self-calibration, open circuit, short circuit, load, 100 groups of customized frequency point 
List scanning 10-point list scanning testing 
Storage device Internal/USB storage device 